Mutagenesis, Vol. 16, No. 2, 133-137,
March 2001
© 2001 UK Environmental Mutagen Society/Oxford University Press
Automatic analysis of slides processed in the Comet assay
Novartis Pharma AG, Preclinical Safety, Toxicology/Pathology, CH-4002 Basel, Switzerland
| Abstract |
|---|
|
|
|---|
In recent years the Comet assay (or single cell gel electrophoresis assay) has been established as a rapid and sensitive method for the detection of DNA damage. For early genotoxicity screening of new chemical entities in industrial toxicology, the Comet assay is more and more used for assessment of the DNA damaging potential of a test compound. In order to increase compound screening throughput, we have established an image analysis system for fully automated measurement of microscope slides processed in the Comet assay. For the comparative investigation various cell types, such as V79 Chinese hamster cells, mouse lymphoma cells and human leukocytes, were treated with several test compounds. Using tail moment as the quantitative parameter for comet formation, we show a very high correlation between our automatic image analysis system and a commercially available, interactive system (Comet Assay II of Perceptive Instruments). The possibility of analyzing 50 samples within 1 day and the high reproducibility of results make automated image processing a powerful tool for automatic analysis of slides processed in the Comet assay.
| Introduction |
|---|
|
|
|---|
The alkaline Comet assay is a well-established genotoxicity test that has been used to detect a broad spectrum of DNA damage with high sensitivity (Fairbairn et al., 1995
Image analysis for quantification of comet intensity was first reported by Olive (Olive, 1989
). While several interactive image analysis systems based on the PC have been established in the meantime, automatic comet analysis has only recently been developed (Böcker et al., 1999
). To facilitate fast screening of large numbers of new compounds in industrial toxicology, complete automation of Comet assay analysis would be a big advantage.
Originally the extent of DNA damage was defined as the head-to-tail ratio by measuring the diameter of the head at a certain staining intensity and dividing by the distance between the head and the end of the comet tail (Olive, 1989
). Various other parameters for comet quantification have since been used (for a review see Ashby et al., 1995; Böcker et al., 1997a, 1999).
We decided to use the tail moment as described by Olive et al. (1990) for the following reasons:
- easy computation through a simple, closed formula (for details see Materials and methods, Image analysis process);
- robustness with respect to degrading fluorescence illumination;
- independence of comet shape;
- well-established parameters which are also used by the interactive comet analysis system.
For fast, sensitive and unbiased analysis of slides we developed an image analysis system with the goal of minimizing user interaction. Except for slide loading and specification of the sample identifier and a starting field at the beginning, no manual steps should be necessary. The resulting system allows fully automatic analysis of 50 comets, which were automatically selected for comet analysis, in <10 min. Through placing two different samples on one slide and by using an electronically driven microscope with a stage capacity of eight slides, up to 16 samples can be analyzed in one run without further user interaction.
This report focuses on the technical aspects of automatic comet analysis using the Leitz MIAS image analysis system. The first step in validating the system was to prove its qualification for automatic comet analysis by establishing doseresponse relationships using the standard mutagens ethylmethane sulfonate (EMS) and 2-aminoanthracene (2-AA) in V79 cells. In a second step, system reliability and usefulness for routine testing were evaluated. The aim was to establish a correlation between the analysis results of the Leitz MIAS system and the results of the Comet Assay II system (Perceptive Instruments), by using test compounds of distinct structural classes and various cell types.
| Materials and methods |
|---|
|
|
|---|
Comet assay
For the Comet assay we used the standard procedure originally described by Singh et al. (1988) with modifications (Klaude et al., 1996
Test compounds
For the first validation step with serial dilutions of standard mutagens in V79 cells the following compounds were used: EMS (Aldrich Chemical Co., USA) without metabolic activation; 2-AA with metabolic activation (10% S9 fraction).
To investigate the reliability of automatic comet analysis with the Leitz MIAS system a comparison with the Perceptive Instruments system was performed, using test compounds that were synthesized in-house as drug candidates. The chemicals were dissolved in either Eagles' minimal essential medium or dimethyl sulfoxide (DMSO). When DMSO was used, the final concentration in the culture medium was 1%. Dose selection was based on cell viability and/or cytotoxicity. In the absence of cytotoxicity all compounds were tested up to precipitating doses or 10 mM or 5 mg/ml, whichever was lower. Solvent control experiments were carried out both in the presence and absence of 10% S9 mix. The S9 mix was prepared using standard conditions (Maron and Ames, 1983
). A liver homogenate was prepared from 79-week-old male Wistar rats [CRL:WI (GLX/BRL/HAN) IGS BR; Charles River, Sulzfeld, Germany]. Five days before being killed the animals were treated with a single i.p. injection of 500 mg/kg Aroclor 1254. The S9 fraction consisted of the supernatant obtained after centrifugation of the liver homogenate for 10 min at 9000 g and was stored at 70 to 85°C until needed. S9 mix, prepared with test tablets (Ames; Boehringer Mannheim, Germany), contained 10% S9 fraction, 4 mM NADP, 5 mM glucose 6-phosphate, 100 mM Na/K phosphate buffer, pH 7.4, 8 mM Mg aspartate and 33 mM KCl.
Cell types used
For our study V79 Chinese hamster and mouse lymphoma L5178Y cells were cultivated under standard conditions. For the tests with human leukocytes peripheral whole blood was collected and treated with the test compounds in RPMI medium.
Cytotoxicity assays
The test substance concentrations for the Comet assay were determined in a cytotoxicity test using a colorimetric method with 3-[4,5-dimethylthiazol-2-yl]-2,5-diphenyl tetrazolium bromide (MTT) (Sigma). The test was performed in 96-well cell culture plates. Cells were treated for 3 h, with or without S9 mix. Forty-eight hours after the start of treatment MTT solution (5 mg/ml in phosphate-buffered saline; Oxoid) was added to the wells. Two hours later the supernatant was removed and DMSO was added (100 µl) to extract the blue color. The plates were then evaluated spectrophotometrically at 570690 nm. The test substance concentration leading to a reduction in staining intensity of
25% of the solvent control was used as the top concentration. For the Comet assay between four and six concentrations of a compound were used. For L5178Y cells Trypan blue dye exclusion was used as a measure of viability. Test compound concentrations resulting in
30% Trypan blue-positive cells were analyzed. Prior to analysis the slides were screened for cells with completely fragmented chromatin (hedgehogs or clouds) (Hartmann and Speit, 1997
). Only samples which showed
15% hedgehogs were evaluated.
Comet analysis
For each concentration of a compound two independent cultures were used. The median tail moment of 50 cells/slide (1 slide/culture) was assessed and the average of the two cultures was calculated. For each compound at least three concentrations per experiment were analyzed. The results were confirmed in a repeat experiment. Generally, a compound was considered positive if the tail moment of at least one concentration with acceptable cytotoxicity was at least twice the corresponding tail moment value of the concurrent solvent control. No formal statistical analysis of the data was carried out. Instead, descriptive parameters, such as an obvious concentration-related increase in the tail moment, or the reproducibility of a specific effect were used for assessment of the data.
Automatic image analysis system
For automatic evaluation of the slides a Leitz MIAS image analyzer (Leitz Messtechnik, Wetzlar, Germany) together with a Leica DM RBE microscope (Leica Microsystems, Wetzlar, Germany) was used. The Leitz MIAS (Modular Image Analysis System) is a computer system using the UNIX-compatible operating system IDRIS with a special hardware architecture for real time image analysis. Morphological transformations on digital images can be efficiently performed by the pipeline architecture of the system, which uses additional video buses besides the standard VME bus. The 68030 CPU is supplemented by boards for image digitization, video display and image storage and dedicated boards for fast morphological gray and binary processing of images.
Image capture was done with a COHU 4912 B/W CCD camera mounted on the microscope equipped with motorized focus, an electronically driven eight slide stage and epifluorescence illumination for green excitation. For comet evaluation a Leica 40x PL APO objective was used.
The software for automatic comet analysis was exclusively developed in-house. It consists of three separate programs: a data input program for specification of the slides to be analyzed; an automatic measurement program; an optional relocation program which enables the operator to retrieve all measured comets for visual inspection.
Image analysis process
The evaluation task consisted of measuring 50 comets/sample. Damaged or otherwise odd-shaped nuclei, particles of cells with completely fragmented chromatin (due to apoptosis or cytotoxic effects) or nuclei that touched the image frame had to be excluded from scoring.
Finding valid nuclei for comet analysis
For image evaluation scanning of non-overlapping, adjacent image frames in a spiral scanning mode, starting at the center of a specimen, was performed. For each stage position a digital 512x512 image with 256 gray values was used. Autofocus was only performed if there were enough `bright' signals, indicating one or more nuclei in the image. This was tested with a simple gray threshold and object analysis of the resulting binary image. If no sufficiently large bright object was present, the stage moved directly to the next image field. Otherwise, autofocus was performed. Due to the high quality of the optical equipment and the fluorescent staining in combination with antifade (Tebbs et al., 1999
), no specific `fluorescence' autofocus had to be developed (Böcker et al., 1997b
). Instead, our standard autofocus algorithm, based on determination of the image with maximum contrast by moving the stage in the z-direction while capturing a series of images and analyzing them, was used, which has been proved to work efficiently for other automated applications (Frieauff et al., 1998
).
After autofocus and image capture, image analysis for a single image field consisted of the following processing steps. An initial thresholding of the original gray image with a fixed threshold which had been determined empirically to optimally segment the nucleus was performed first. After some morphological opening and closing for removal of small bright noise particles (including comet tail particles!), as well as closing small gaps of nucleus segmentation, an individual analysis of the segmented binary objects was carried out. For details on these morphological operations see Serra (1986). Considering metric parameters such as area and shape, as well as features like `touching the image frame' or `distance to other large, bright objects', each segmented object was classified either as `valid nucleus for comet analysis' or as `artifact to be rejected from further consideration'. Burst or overlapping nuclei as well as other debris were essentially eliminated at this stage. The coordinates of each accepted valid nucleus were stored for comet analysis at a later point in time. If no valid nucleus passed the metric analysis the current image field analysis was terminated and the next field was selected.
Comet analysis
After 50 valid nuclei had been detected the nucleus finding process was stopped for the current sample and the stage automatically returned to the first valid nucleus. For comet analysis each nucleus which had been previously accepted was positioned in the center of the image field. This guaranteed unbiased analysis of the comets without the influence of varying shading effects in different regions of the image field (due to uneven fluorescence illumination), as well as the complete comet within the current image.
An autofocus was again performed, but this time on a rectangular sub-range around the comet in order to maximize the contrast of the desired object alone. After image grabbing, threshold determination for comet segmentation was performed using a similar procedure to that described by Böcker et al. (1999). The optimal threshold was determined by successively decreasing the threshold from a starting value and counting the object number in the resulting binary image:
OBJ = Obj_Nr(Binary_Image(Current_Thresh)) Obj_Nr(Binary_Image(Previous_Thresh))
where Binary_Image(Current_Thresh) denotes the binary image resulting from thresholding after having decreased Previous_Thresh, while Binary_Image(Previous_Thresh) was the binary image resulting from thresholding with the old threshold.
If OBJ exceeded a certain limit (200 particles in our case) we considered the computed threshold plus a constant, empirically determined offset to be the best threshold for comet segmentation.
The rationale of this approach was that through decreasing the threshold one approached the dark background, increasing noise particle segmentation, which led to a drastic increase in the number of objects in the image. So, before erroneously segmenting additional noise objects, the small comet tail particles could still be included. The gradient-like formula for OBJ yielded a more robust and independent segmentation result of the actual number of comets in the image field as compared with a direct limit on the number of binary objects in the thresholded image, as described by Böcker et al. (1999).
After ensuring that the minimum distance of the centered nucleus to the next large, bright object (another nucleus or artifact) was exceeded, an analysis of the segmented objects characterizing the comet tail particles was performed. They should be located within a certain rectangular frame with the nucleus at one end and the distance between individual particles was not allowed to exceed a certain limit in order to discard bright particles not belonging to the comet.
The tail moment (TM) for the comet was computed by the following formula, a modification of the Olive tail moment (Olive et al., 1990
):
![]() |
![]() |
For retrieval and visual inspection of the measured comets by the relocation program their stage coordinates were stored together with the individual tail moments.
After all valid nuclei had been relocated and analyzed for comet tail moment the median over the resulting 50 individual tail moment values was computed to yield the overall descriptive parameter for the sample.
Experiments for the validation of automated analysis versus manual scoring
For validation of the reliability and usefulness of automatic image analysis for the Comet assay as a screening tool, automatic analysis was carried out in parallel with manual analysis (Comet Assay II; Perceptive Instruments). The first series of experiments consisted of treatment of V79 cells with various concentrations of standard mutagens. Furthermore, reproducibility of the test results was addressed by repeated analysis of slides with the automatic system. In a next step, in-house developed compounds with unknown genotoxic potential were used. These tests were conduced with V79 cells, L5178Y cells or human leukocytes. The aim was to clarify whether it was possible to clearly identify positive, weakly positive or negative effects with the MIAS system with the same sensitivity as the Perceptive Instruments system.
| Results |
|---|
|
|
|---|
Establishing doseresponse curves with the different image analysis systems
For the first validation step V79 cells were treated with the direct acting mutagen EMS in the absence of S9 mix at 40, 100, 200, 330 or 400 µg/ml for 3 h. 2-AA was used with metabolic activation at concentrations of 17, 20, 21, 22 or 23 µg/ml. Fifty comets per sample were analyzed, using both the Leitz MIAS automatic system and the Perceptive Instruments Comet Assay II system.
The results for EMS treatment are shown in Figure 1A and B
. For both systems a clear concentration-dependent increase in tail moment was found. The reproducibility of the results (repeated analysis of the same slides after 7 days) was very good. In the case of automatic Leitz MIAS only a slight decrease in tail moment was seen at high concentrations of EMS. This difference may be due to the prolonged storage, however, this does not change the overall picture. The results of two individuals scoring slides with the Perceptive Instruments system differ only marginally at higher concentrations of EMS and may reflect the influence of nucleus selection on the resulting tail moment values.
|
Figure 2
|
Automatic versus interactive comet analysis
Table I
|
|
The 6-fold increase in tail moment measured with MIAS as compared with the Perceptive Instruments values can be explained by the different frame grabbers of the two systems, microscopes with different optical components and, possibly, different implementations of tail moment computation.
Relocation of measured cells
Relocation of the analyzed comets showed that erroneous analysis of invalid comets (touching nuclei, clouds and otherwise misshapen objects) was no problem since computation of the median of tail moments meant `ignoring' those few outliers which had no influence on the overall tail moment median of a sample.
Evaluation time
The interactive analysis of one sample (50 comets) with the Comet Assay II system (Perceptive Instruments) took ~10 min. As for automatic image analysis, the evaluation time depended upon the distribution density of nuclei on the slide.
The optimum comet density was one per image field on average. If there were more than four nuclei per field the minimum distance constraint for valid nuclei led to skipping of all comets and less than one comet per four image fields on average implied that a high number of `empty' image fields had to be checked.
With the MIAS system the typical evaluation time for a standard sample with average cell density was 8 min for 50 comets (50100 image fields to be analyzed). Sample specification and automatic analysis of 16 samples (i.e. eight slides) usually took <3 h, which made the time needed for automatic comet analysis comparable to interactive scoring.
| Discussion and conclusion |
|---|
|
|
|---|
The Comet assay is used in our laboratory as part of our screening test battery for the genotoxicity assessment of new chemical entities. The aim of this study was to automate slide analysis for faster compound throughput, equivalent to screening more compounds in the same time.
We developed a computerized image analysis system for fully automated analysis of samples processed in the Comet assay. Our Leitz MIAS system can process up to 16 samples (eight slides, 2 samples/slide) without any user interaction. System performance proved to be independent of the cell type (V79 Chinese hamster cells, L5178Y mouse lymphoma cells and human leukocytes).
A series of tests for comparing interactive scoring with automatic image analysis showed that there was an excellent concordance between the results obtained with the two methods. Using the same criteria for a positive effect, the conclusions were the same in all experiments, irrespective of the scoring method and cell type used. Thus, automatic image analysis showed a high degree of reliability and well-known mutagens were easily detected. Weakly positive effects were detected with similar sensitivity by automatic image analysis as compared with manual comet scoring.
Since slide reading by image analysis is a fully automatic process, 50 samples/day can be easily analyzed. Little time is required for slide loading and sample specification. In summary, automatic image analysis using Leitz MIAS has proved to be an excellent system for fast, reliable and objective comet assay analysis for various cell types. The software package, consisting of programs for slide-specific information entry, automatic detection and optional relocation of the measured comets, has been shown to be reliable, efficient and user friendly. Since the Leitz MIAS system is no longer commercially available and due to faster PC image processing as well as more convenient PC software development tools, we recently decided to develop a robot-based, fully automated image analysis system based on the PC. This system, which is intended to be commercially available, will be the platform for a variety of automatic image analysis applications in toxicology.
Based on the results presented, we decided to replace interactive slide analysis by automatic slide reading in all our Comet assay experiments, conducted as screening tests in the genotoxicity assessment of potential new pharmaceuticals. This decision has led to a doubling in the throughput of compounds screened per technician, as compared with interactive scoring.
| Acknowledgments |
|---|
We gratefully acknowledge the technical assistance of C.Blaser and M.Schneider in the laboratory.
| Notes |
|---|
1 To whom correspondence should be addressed. Tel: +41 61 3241595; Fax: +41 61 3241274; Email: wilfried.frieauff{at}pharma.novartis.com
| References |
|---|
|
|
|---|
-
Anderson,D., Yu,T.-W. and McGregor,D.B. (1998) Comet assay responses as indicators of carcinogen exposure. Mutagenesis, 539555.
Ashby,J., Tinwell,H., Lefevre,P.A. and Browne,M.A. (1995) The single cell gel electrophoresis assay for induced DNA damage (comet assay): measurement of tail length and moment. Mutagenesis, 10, 8590.
Böcker,W., Bauch,T., Müller,W.-U. and Streffer,C. (1997a) Technical report: Image analysis of comet assay measurements. Int. J. Radiat. Biol., 72, 449460.[Web of Science][Medline]
Böcker,W., Rolf,W., Müller,W.-U. and Streffer,C. (1997b) A fast autofocus unit for fluorescence microscopy. Phys. Med. Biol., 42, 19811992.[Web of Science][Medline]
Böcker,W., Rolf,W., Bauch,T., Müller,W.-U. and Streffer,C. (1999) Automated Comet assay analysis. Cytometry, 35, 134144.[Web of Science][Medline]
Fairbairn,D.W., Olive,P.L. and O'Neill,K.L. (1995) The comet assay: a comprehensive review. Mutat. Res., 339, 3759.[Web of Science][Medline]
Frieauff,W., Pötter-Locher,F., Cordier,A. and Suter,W. (1998) Automatic analysis of the in vitro micronucleus test on V79 cells. Mutat. Res., 413, 5768.[Web of Science][Medline]
Hartmann,A. and Speit,G. (1997) The contribution of cytotoxicity to DNA-effects in the single cell gel test (comet assay). Toxicol. Lett., 90, 183188.[Web of Science][Medline]
Hartmann,A., Pfuhler,S., Dennog,C., Germadnik,D., Pilger,A. and Speit,G. (1998) Exercise-induced DNA effects in human leukocytes are not accompanied by increased formation of 8-hydroxy-2'-deoxyguanosine or induction of micronuclei. Free Radic. Biol. Med., 24, 245251.[Web of Science][Medline]
Hartmann,A., Elhajouji,A., Poetter,F., Martus,H.-J., Fjällmann,A., Frieauff,W. and Suter,W. (2000) Use of the alkaline comet assay for industrial genotoxicity screening: comparative investigation with the micronucleus test. Environ. Mol. Mutagen., submitted for publication.
Klaude,M., Ericson,S., Nygren,J. and Ahnström,G. (1996) The comet assay: mechanisms and technical considerations. Mutat. Res., 363, 8996.[Web of Science][Medline]
Maron,D.M. and Ames,B.N. (1983) Revised methods for the Salmonella mutagenicity test. Mutat. Res., 113, 173215.[Web of Science][Medline]
Olive,P.L. (1989) Cell proliferation as a requirement for development of the contact effect in Chinese hamster V79 spheroids. Radiat. Res., 117, 7992.[Web of Science][Medline]
Olive,P.L., Banath,J.P. and Durand,R.E. (1990) Detection of etoposide resistance by measuring DNA damage in individual Chinese hamster cells. J. Natl Cancer Inst., 82, 779783.
Rojas,E., Lopez,M.C. and Valverde,M. (1999) Single cell electrophoresis: methodology and applications. J. Chromatogr., 722B, 225254.
Serra,J. (1986) Introduction to mathematical morphology. Comput. Vis. Graph. Image Process., 35, 283305.
Singh,N.P., McCoy,M.T., Tice,R.R. and Schneider,E.L. (1988) A simple technique for quantitation of low levels of DNA damage in individual cells. Exp. Cell Res., 175, 184191.[Web of Science][Medline]
Tebbs,R.S., Pedersen,R.A., Cleaver,J.E. and Hartmann,A. (1999) Modification of the comet assay for the detection of DNA strand breaks in extremely small tissue samples. Mutagenesis, 14, 437438.
Tice,R.R., Agurell,E., Anderson,D., Burlinson,B., Hartmann,A., Kobayashi,H., Miyamae,Y., Rojas,E., Ryu,J.-C. and Sasaki,Y.F. (2000) The single cell gel/comet assay: guidelines for in vitro and in vivo genetic toxicology testing. Environ. Mol. Mutagen., 35, 206221.[Web of Science][Medline]
Received on August 1, 2000; accepted on October 16, 2000.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
M. D. Fellows and M. R. O'Donovan Cytotoxicity in cultured mammalian cells is a function of the method used to estimate it Mutagenesis, July 1, 2007; 22(4): 275 - 280. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Witte, U. Plappert, H. de Wall, and A. Hartmann Genetic Toxicity Assessment: Employing the Best Science for Human Safety Evaluation Part III: The Comet Assay as an Alternative to In Vitro Clastogenicity Tests for Early Drug Candidate Selection Toxicol. Sci., May 1, 2007; 97(1): 21 - 26. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Hartmann, M. Schumacher, U. Plappert-Helbig, P. Lowe, W. Suter, and L. Mueller Use of the alkaline in vivo Comet assay for mechanistic genotoxicity investigations Mutagenesis, January 1, 2004; 19(1): 51 - 59. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Kiskinis, W. Suter, and A. Hartmann High throughput Comet assay using 96-well plates Mutagenesis, January 1, 2002; 17(1): 37 - 43. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||